User Contributed MET/CAL PROCEDURE ================================================================================ INSTRUMENT: HEWLETT-PACKARD 1742A: PART 4 DATE: 09-Jun-82 AUTHOR: User Contributed REVISION: V1.0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 24 NUMBER OF LINES: 122 CONFIGURATION: 66 =============================================================================== STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK+ R N P J S M A L T W 1.002 ASK- P F 1.003 HEAD {DEFLECTION FACTOR} 1.004 DISP Connect the pulsehead to the channel A input and a 50 1.004 DISP ohm BNC cable to the MAIN EXT TRIGGER input... 1.005 SET A COUPLING: GND 1.005 SET DISPLAY MODE: A 1.006 DISP Use the A POSITION control to set the trace on the 1.006 DISP bottom graticule of the display... 1.007 SET A COUPLING: DC 1.008 MEM2 ={Channel A} 1.009 5001 20 160VP 3% 1.010 JMP 11.002 2.001 5001 10 80VP 3% 2.002 JMP 11.002 3.001 5001 5 40VP 3% 3.002 JMP 11.002 4.001 5001 2 16VP 3% 4.002 JMP 11.002 5.001 5001 1 8VP 3% 5.002 JMP 11.002 6.001 5001 .5 4VP 3% 6.002 JMP 11.002 7.001 5001 .2 1.6VP 3% 7.002 JMP 11.002 8.001 5001 .1 .8VP 3% 8.002 JMP 11.002 9.001 5001 .05 .4VP 3% 9.002 JMP 11.002 10.001 5001 .02 .16VP 3% 10.002 JMP 11.002 11.001 JMP 11.007 11.002 SET A VOLTS/DIV: .2V 11.003 5001 1VP S 11.004 DISP {Adjust A3R65 (composite gain) on Vertical Preamplifier} 11.004 DISP Assy board for 5 divisions[DRAW33,55,T] 11.005 5001 * S 11.006 JMP 1.009 11.007 SET DISPLAY MODE: B 11.008 MEM2 ={Channel B} 11.009 DISP Disconnect pulse head from Channel A and connect to 11.009 DISP Channel B... 11.010 SET B COUPLING: GND 11.011 DISP Use the B POSITION control to position the trace on the 11.011 DISP bottom graticule of the display... 11.012 SET B COUPLING: DC 11.013 5001 20 160VP 3% 11.014 JMP 21.003 12.001 5001 10 80VP 3% 12.002 JMP 21.003 13.001 5001 5 40VP 3% 13.002 JMP 21.003 14.001 5001 2 16VP 3% 14.002 JMP 21.003 15.001 5001 1 8VP 3% 15.002 JMP 21.003 16.001 5001 .5 4VP 3% 16.002 JMP 21.003 17.001 5001 .2 1.6VP 3% 17.002 JMP 21.003 18.001 5001 .1 .8VP 3% 18.002 JMP 21.003 19.001 5001 .05 .4VP 3% 19.002 JMP 21.003 20.001 5001 .02 .16VP 3% 20.002 JMP 21.003 21.001 5001 * S 21.002 JMP 21.009 21.003 SET B VOLTS/DIV: .2V 21.004 5001 1VP S 21.005 DISP {Adjust A3R46 (Channel B GAIN)} for 5 divisions of 21.005 DISP deflection [DRAW33,55,T] 21.006 5001 * S 21.007 JMP 11.013 21.008 CON 0 21.009 DISP Disconnect all connections to the Unit Under Test... 21.010 HEAD {CALIBRATOR VERIFICATION} 21.011 SET A VOLTS/DIV: .2V 21.011 SET MAIN TIME/DIV: .5mS 21.012 SET DISPLAY MODE: A 21.013 DISP Connect the CAL 1V output to the input of the comparator 21.013 DISP head marked FROM DUT CALIBRATOR. Connect the CHOPPED 21.013 DISP OUTPUT of the comparator to channel A INPUT. 21.014 MESS Set the comparator head on Int. termination 'OPEN' and 21.014 MESS depress 'AUTO'. Adjust stimulus for waveform coincidence 21.015 5001 1VP 1% -D CO 22.001 DISP Disconnect the comparator head and connect the CAL 1V 22.001 DISP output to Channel A input. 22.002 SET A VOLTS/DIV: .1V 22.002 SET MAIN TIME/DIV: .05uS 22.003 MESS Rotate variable 'CAL' to obtain 6 divisions of display 22.004 EVAL {Is the rise time < .1uS ?} 23.001 HEAD {ASTIGMATISM AND FOCUS ADJUSTMENT} 23.002 SET MAIN TIME/DIV: 1S 23.003 DISP TIME/DIV VERNIER: FULLY COUNTER-CLOCKWISE[DRAW29,75,F] 23.004 DISP BEAM INTENSITY: BARELY VISIBLE SPOT[DRAW57,60,F] 23.005 MESS While spot slowly moves across CRT, adjust FOCUS on 23.005 MESS panel and ASTIGMATISM on rear panel for smallest and 23.005 MESS best defined spot 23.006 EVAL {Can a well defined spot be obtained ?} 24.001 DISP Disconnect all connections to the Unit Under Test... END